Abstract

Citation: Clin Surg. 2018;3(1):2125.Research Article | Open Access

Destabilization of the Comminuted Clavicle Shaft Fracture due to Breakage of the Titanium Locking Plate: Biomechanical Analysis

Grzegorz Szczęsny and Zbigniew Kowalewski

Department of Orthopedic Surgery and Traumatology, Medical University, Poland
Department of Experimental Mechanics, IPPT, Polish Academy of Sciences, Poland

*Correspondance to: Grzegorz Szcz?sny 

 PDF  Full Text DOI: 10.25107/2474-1647.2125

Abstract

Background: Implant breaking destabilizing the fracture may be caused by its structural defect,inappropriate use, or massive overloads. We present the patient, who broke the plate stabilizing comminuted clavicular fracture at the sixth postoperative week due to the, primarily reported, simple loading of the limb that should not exceeded admissible limits.
Methods: Macroscopic examination excluded its manufacturing defect of an explant, but revealed areas of brittle and fatigue fractures, as well as anterior and inferior bending of its edges. Mechanical analysis proved sufficient material durability. Those findings clearly showed that the plate broke in consequence of repetitive overloads exceeding its durability. Patient, when acquitted with those results, confessed that the plate broke, while fishing. Moreover, he reported an episode of alcohol overuse that could contribute to implant failure bringing the risk of additional injury.Results: Biomechanical analysis confirmed that mechanical loads throwing the spinner exceed values adequate for plate breaking. Thus, the real circumstances of the complication seem to be unveiled pointing out to facts that some patients disobey postoperative recommendations to avoid overloading of stabilized extremity, thus participating actively in this type of complications, that they dissemble the real circumstances of the complications, and that they may exhibit risky and irrational behavior.Conclusion: Basing on the presented case we concluded that in some cases plaster cast immobilization supporting stabilized fracture may be beneficial precluding banned activities and preventing from secondary injuries reducing the risk of postoperative complications.

Keywords

Clavicle fracture; Implant failure; Mechanical overloads; Postoperative recommendations disobey; Insubordinate patient

Cite the article

Szcz?sny G, Kowalewski Z. Destabilization of the Comminuted Clavicle Shaft Fracture due to Breakage of the Titanium Locking Plate: Biomechanical Analysis. Clin Surg. 2018; 3: 2125.

Search Our Journal

Journal Indexed In

Articles in PubMed

RAF Kinase Inhibitory Protein Expression and Phosphorylation Profiles in Oral Cancers
 PubMed  PMC  PDF  Full Text
Mesh Sprayer Device with Liquefied Mesh Delivery System: Proposed Alternative for Currently Available Meshes in Hernia Repair and Supplement to Abdominal Closure
 PubMed  PMC  PDF  Full Text
View More...

Articles with Grants

Association of Methylation of RASSF1A between Breast Cancer and Benign Breast Lesions: A Meta-Analysis
 Abstract  PDF  Full Text
Chronic Mesenteric Volvulus with Superior Mesenteric Venous Thrombosis: A Case Report
 Abstract  PDF  Full Text
View More...